Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

LC Multiresidue Pesticide Standard #2 on Raptor ARC-18 by LC-MS/MS

LC_FF0544
PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Aldicarb sulfoxide1.65207.089.0132.0
2.Butoxycarboxim1.67223.0106.0166.0
3.Aldicarb sulfone1.71240.0148.086.0
4.Oxamyl1.78237.072.090.0
5.Methomyl1.90163.0106.088.0
6.Butocarboxim2.68213.0156.0116.0
7.Aldicarb2.71213.189.1116.1
8.Thiophanate-methyl2.87343.0151.093.0
PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
9.Thidiazuron2.92221.0101.993.9
10.Tebuthiuron3.07229.0172.0116.0
11.Ethiofencarb3.19226.1107.0164.0
12.Methabenzthiazuron3.46222.0165.0150.0
13.Methiocarb3.95226.0169.0121.0
14.Alanycarb5.05400.0238.2254.1
15.Benfuracarb5.76411.1195.0190.0
16.Furathiocarb5.81383.2194.9252.0
ColumnRaptor ARC-18 (cat.# 9314A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:50 °C
Standard/SampleLC multiresidue pesticide standard #2 (cat.# 31973)
Diluent:Water
Conc.:20 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water + 2 mM ammonium formate + 0.2% formic acid
B:Methanol + 2 mM ammonium formate + 0.2% formic acid
Time (min)Flow (mL/min)%A%B
0.000.4955
2.000.44060
4.000.42575
6.000.40100
7.000.40100
7.010.4955
9.500.4955
DetectorWaters Xevo TQ-S
Ion Source:Waters Zspray™ ESI
Ion Mode:ESI+
Mode:MRM
InstrumentWaters
Notes• Autosampler temperature = 5 °C
• The reference standard is also available as part of Restek’s 200+ compound LC multiresidue pesticide kit (cat.# 31971).