Peaks | tR (min) | Precursor Ion | Product Ion 1 | Product Ion 2 | |
---|---|---|---|---|---|
1. | Aldicarb sulfoxide | 1.65 | 207.0 | 89.0 | 132.0 |
2. | Butoxycarboxim | 1.67 | 223.0 | 106.0 | 166.0 |
3. | Aldicarb sulfone | 1.71 | 240.0 | 148.0 | 86.0 |
4. | Oxamyl | 1.78 | 237.0 | 72.0 | 90.0 |
5. | Methomyl | 1.90 | 163.0 | 106.0 | 88.0 |
6. | Butocarboxim | 2.68 | 213.0 | 156.0 | 116.0 |
7. | Aldicarb | 2.71 | 213.1 | 89.1 | 116.1 |
8. | Thiophanate-methyl | 2.87 | 343.0 | 151.0 | 93.0 |
Peaks | tR (min) | Precursor Ion | Product Ion 1 | Product Ion 2 | |
---|---|---|---|---|---|
9. | Thidiazuron | 2.92 | 221.0 | 101.9 | 93.9 |
10. | Tebuthiuron | 3.07 | 229.0 | 172.0 | 116.0 |
11. | Ethiofencarb | 3.19 | 226.1 | 107.0 | 164.0 |
12. | Methabenzthiazuron | 3.46 | 222.0 | 165.0 | 150.0 |
13. | Methiocarb | 3.95 | 226.0 | 169.0 | 121.0 |
14. | Alanycarb | 5.05 | 400.0 | 238.2 | 254.1 |
15. | Benfuracarb | 5.76 | 411.1 | 195.0 | 190.0 |
16. | Furathiocarb | 5.81 | 383.2 | 194.9 | 252.0 |
Column | Raptor ARC-18 (cat.# 9314A12) | ||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 2.1 mm ID | ||||||||||||||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||||||||||||||
Temp.: | 50 °C | ||||||||||||||||||||||||||||||||
Standard/Sample | LC multiresidue pesticide standard #2 (cat.# 31973) | ||||||||||||||||||||||||||||||||
Diluent: | Water | ||||||||||||||||||||||||||||||||
Conc.: | 20 ng/mL | ||||||||||||||||||||||||||||||||
Inj. Vol.: | 5 µL | ||||||||||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||||||||||
A: | Water + 2 mM ammonium formate + 0.2% formic acid | ||||||||||||||||||||||||||||||||
B: | Methanol + 2 mM ammonium formate + 0.2% formic acid | ||||||||||||||||||||||||||||||||
|
Detector | Waters Xevo TQ-S |
---|---|
Ion Source: | Waters Zspray™ ESI |
Ion Mode: | ESI+ |
Mode: | MRM |
Instrument | Waters |
Notes | Autosampler Temperature = 5 °C |