Peaks | tR (min) | Precursor Ion | Product Ion 1 | Product Ion 2 | |
---|---|---|---|---|---|
1. | trans-Piceid* | 0.78 | 389.1 | 227.0 | 142.9 |
2. | cis-Piceid* | 0.94 | 389.1 | 227.0 | 142.9 |
3. | trans-Resveratrol | 1.06 | 227.0 | 184.9 | 142.6 |
4. | cis-Resveratrol | 1.17 | 227.0 | 184.9 | 142.6 |
Column | Raptor ARC-18 (cat.# 9314552) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 50 mm x 2.1 mm ID | ||||||||||||||||||||
Particle Size: | 5 µm | ||||||||||||||||||||
Temp.: | 40 °C | ||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||
Diluent: | Water | ||||||||||||||||||||
Conc.: | 25 ng/mL | ||||||||||||||||||||
Inj. Vol.: | 5 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | Water + 0.1% formic acid | ||||||||||||||||||||
B: | Acetonitrile + 0.1% formic acid | ||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Mode: | ESI- |
Instrument | HPLC |