Peaks | tR (min) | Conc. (ng/mL) | Precursor Ion | Product Ion | Product Ion | |
---|---|---|---|---|---|---|
1. | Acesulfame-K | 0.83 | 100 | 162.04 | 82.07 | 77.97 |
2. | Cyclamate | 1.07 | 100 | 178.14 | 79.99 | 96.00 |
3. | Saccharin | 1.17 | 100 | 182.02 | 106.01 | 61.97 |
4. | Sucralose | 1.45 | 100 | 441.23 | 395.17 | 359.21 |
5. | Aspartame | 1.50 | 100 | 293.27 | 261.19 | 200.12 |
6. | Alitame | 1.58 | 100 | 330.30 | 312.23 | 295.19 |
7. | Dulcin | 1.83 | 100 | 181.13 | 138.60 | 107.64 |
8. | Neohesperidin dihydrochalcone | 2.17 | 100 | 611.40 | 303.21 | 166.13 |
9. | Stevioside | 2.31 | 100 | 803.58 | 641.52 | 479.45 |
10. | Neotame | 2.45 | 100 | 377.34 | 200.15 | 345.30 |
11. | Glycyrrhizic acid | 2.83 | 100 | 821.67 | 351.19 | 113.07 |
Column | Raptor Biphenyl (cat.# 9309252) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 50 mm x 2.1 mm ID | ||||||||||||||||||||
Particle Size: | 1.8 µm | ||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||
Temp.: | 40 °C | ||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||
Diluent: | Water | ||||||||||||||||||||
Conc.: | 100 ng/mL | ||||||||||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | Water, 0.1% formic acid | ||||||||||||||||||||
B: | Acetonitrile, 0.1% formic acid | ||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Mode: | MRM |
Instrument | UHPLC |
Notes | Dulcin was analyzed in ESI+ mode. All other compounds were analyzed in ESI- mode. |