Peaks | tR (min) | Precursor | Product 1 | Product 2 | Product 3 | |
---|---|---|---|---|---|---|
1. | Monocrotaline | 0.677 | 326.20 | 120.3 | 121.30 | 94.25 |
2. | Incidine | 1.111 | 300.20 | 94.15 | 138.20 | 156.36 |
3. | Heliotrine | 1.763 | 314.30 | 138.20 | 156.20 | 120.20 |
4. | Heliotrine N-oxide | 1.895 | 330.25 | 172.25 | 138.20 | |
5. | Senecionine | 2.240 | 336.20 | 120.30 | 138.30 | |
6. | Senecionine N-oxide | 2.410 | 352.35 | 94.20 | 120.30 | 118.25 |
7. | Echimidine | 2.524 | 398.30 | 120.25 | 220.15 | |
8. | Echimidine N-oxide | 2.528 | 414.25 | 396.30 | 352.20 | 254.25 |
9. | Lasiocarpine | 3.028 | 412.20 | 120.20 | 220.15 | 336.25 |
10. | Lasiocarpine N-oxide | 3.260 | 428.25 | 254.25 | 352.25 | 120.30 |
Column | Raptor ARC-18 (cat.# 9314A52) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 50 mm x 2.1 mm ID | ||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||
Temp.: | 25 °C | ||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||
Diluent: | 0.1% Formic acid in water | ||||||||||||||||||||
Conc.: | 10 ng/mL | ||||||||||||||||||||
Inj. Vol.: | 1 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | 0.1% Formic acid in water | ||||||||||||||||||||
B: | 0.1% Formic acid in acetonitrile | ||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Mode: | ESI+ |
Instrument | UHPLC |