Peaks | tR (min) | Conc. (µg/g) | Precursor Ion | Product Ion 1 | Product Ion 2 | Product Ion 3 | Polarity | |
---|---|---|---|---|---|---|---|---|
1. | Orientin | 1.469 | 0.1 | 449 | 329.1 | 299.1 | 431 | + |
2. | Vitexin | 1.827 | 0.1 | 433 | 313.1 | 283 | 415.1 | + |
3. | Rutin | 1.895 | 1 | 609.1 | 300.1 | 301 | 271 | - |
4. | Quercetin-3-D-glycopyranoside | 2.01 | 0.1 | 463.1 | 300 | 301 | 271 | - |
5. | Isovitexin | 2.024 | 0.1 | 433.4 | 283 | 313.1 | 337 | + |
6. | Luteolin-7-O-glucuronide | 2.221 | 0.1 | 463 | 287.1 | 153.2 | 135.1 | + |
7. | Fisetin | 2.435 | 1 | 285 | 135 | 121 | 163 | - |
8. | Apigenin-7-O-glucuronide | 2.664 | 0.1 | 447 | 271.1 | 153.2 | 119.1 | + |
9. | Quercetin | 2.753 | 1 | 301 | 151 | 179 | 107 | - |
10. | Luteolin | 2.919 | 1 | 285 | 133 | 151 | 175 | - |
11. | Kaempferol | 3.137 | 1 | 285 | 92.9 | 239 | 185 | - |
12. | Apigenin | 3.266 | 0.1 | 271 | 153 | 119 | 69.1 | + |
13. | Silymarin | 3.322 | 1 | 481.1 | 301 | 125.1 | 152 | - |
14. | Baicalin | 3.363 | 0.1 | 447.1 | 271.1 | 123.1 | 253.1 | + |
15. | Chrysin | 4.141 | 1 | 253 | 143.2 | 63.1 | 209 | - |
16. | Wogonin | 4.220 | 0.1 | 285 | 270.1 | 151.2 | 179 | + |
17. | Cannflavin B | 4.321 | 0.1 | 369.1 | 313 | 298 | 165 | + |
18. | Cannflavin A | 4.620 | 0.1 | 437.1 | 313.1 | 165 | 298.1 | + |
19. | B-sitosterol | 5.472 | 1 | 397.4 | 161.1 | 135.1 | 147.2 | + |
Column | Raptor Biphenyl (cat.# 9309A12) | ||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 2.1 mm ID | ||||||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||||||
Guard Column: | Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252) | ||||||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||||||
Diluent: | Water:methanol (60:40) | ||||||||||||||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||
A: | Water, 0.1% formic acid | ||||||||||||||||||||||||
B: | Methanol, 0.1% formic acid | ||||||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Source: | Electrospray |
Ion Mode: | ESI+/ESI- |
Mode: | MRM |
Instrument | UHPLC |