Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | tR (min) | Conc. (µg/mL) | |
---|---|---|---|
1. | Cannabidivarinic acid (CBDVA) | 1.24 | 50 |
2. | Cannabidivarin (CBDV) | 1.375 | 50 |
3. | Cannabidiolic acid (CBDA) | 1.743 | 50 |
4. | Cannabigerolic acid (CBGA) | 1.858 | 50 |
5. | Cannabigerol (CBG) | 1.963 | 50 |
6. | Cannabidiol (CBD) | 2.065 | 50 |
7. | Tetrahydrocannabivarin (THCV) | 2.237 | 50 |
8. | Tetrahydrocannabivarinic acid (THCVA) | 2.848 | 50 |
Peaks | tR (min) | Conc. (µg/mL) | |
---|---|---|---|
9. | Cannabinol (CBN) | 3.077 | 50 |
10. | Cannabinolic acid (CBNA) | 3.637 | 50 |
11. | Δ9-Tetrahydrocannabinol (Δ9-THC) | 3.882 | 50 |
12. | Δ8-Tetrahydrocannabinol (Δ8-THC) | 4.014 | 50 |
13. | Cannabicyclol (CBL) | 4.621 | 50 |
14. | Cannabichromene (CBC) | 4.885 | 50 |
15. | Tetrahydrocannabinolic acid A (THCA-A) | 5.134 | 50 |
16. | Cannabichromenic acid (CBCA) | 5.75 | 50 |
Column | Raptor ARC-18 (cat.# 9314A6E) | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 150 mm x 3 mm ID | ||||||||||||
Particle Size: | 2.7 µm | ||||||||||||
Pore Size: | 90 Å | ||||||||||||
Temp.: | 30 °C | ||||||||||||
Standard/Sample | |||||||||||||
Diluent: | 25:75 Water:acetonitrile | ||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||
Mobile Phase | |||||||||||||
A: | Water, 5 mM ammonium formate, 0.1% formic acid | ||||||||||||
B: | Acetonitrile, 0.1% formic acid | ||||||||||||
|
Detector | UV/Vis @ 228 nm |
---|---|
Instrument | HPLC |