Peaks | tR (min) | |
---|---|---|
1. | Cannabidiol (CBD) | 3.259 |
Column | Raptor ARC-18 (cat.# 9314A65) | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 150 mm x 4.6 mm ID | ||||||||||||
Particle Size: | 2.7 µm | ||||||||||||
Pore Size: | 90 Å | ||||||||||||
Guard Column: | Raptor ARC-18 EXP guard column cartridge 5 mm, 4.6 mm ID, 2.7 µm (cat.# 9314A0250) | ||||||||||||
Temp.: | 30 °C | ||||||||||||
Standard/Sample | |||||||||||||
Conc.: | Endogenous concentration | ||||||||||||
Inj. Vol.: | 5 µL | ||||||||||||
Mobile Phase | |||||||||||||
A: | Water, 5 mM ammonium formate, 0.1% formic acid | ||||||||||||
B: | Acetonitrile, 0.1% formic acid | ||||||||||||
|
Detector | UV/VIS @ 228 nm |
---|---|
Sample Preparation | Beverage was sonicated for 30 min to remove carbonation. 10 mL of beverage was added to a 50 mL centrifuge tube. 10 mL of acetonitrile was added along with an EN Q-sep QuEChERS salt packet (cat.# 25849), and the sample was shaken by hand for 1 minute and then vortexed for 30 seconds. The sample was transferred to a shaker table and shaken for 10 minutes followed by centrifugation at 4200 rpm for 10 minutes. An aliquot was transferred to an autosampler vial prior to analysis. |