Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Trifluoroacetic Acid (TFA) on Raptor Polar X

LC_FF0621
PeakstR (min)Precursor IonProduct Ion
1.Trifluoroacetic acid (TFA)1.29113.0069.05
ColumnRaptor Polar X (cat.# 9311A32)
Dimensions:30 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:40 °C
Standard/Sample
Diluent:15:85 Mobile phase A:mobile phase B
Conc.:200 ng/L
Inj. Vol.:3 µL
Mobile Phase
A:Water, 10 mM ammonium formate, 0.05% formic acid
B:80:20 Acetonitrile:methanol, 0.05% formic acid
Time (min)Flow (mL/min)%A%B
0.000.51585
20.51585
DetectorShimadzu 8060 MS/MS
Ion Source:Electrospray
Ion Mode:ESI-
Mode:MRM
InstrumentShimadzu Nexera X2
Sample PreparationThe sample was aliquoted into a 2.0 mL, 9 mm screw-thread polypropylene vial (cat.# 23242) and capped with a 9 mm solid-top polyethylene cap (cat.# 23244).