Peaks | tR (min) | Conc. (ng/mL) | Precursor 1 | Product 1 | Product 2 | Precursor 2 | Product 1 | ESI Polarity | |
---|---|---|---|---|---|---|---|---|---|
1. | BAC-C8 | 1.10 | 50 | 248.0 | 91.0 | 156.2 | - | - | + |
2. | BAC-C10 | 2.12 | 50 | 276.0 | 91.1 | 184.2 | - | - | + |
3. | DDAC-C8 | 2.67 | 50 | 270.3 | 158.2 | - | - | - | + |
4. | BAC-C12 | 3.16 | 50 | 304.0 | 91.1 | 212.2 | - | - | + |
5. | Chlorate | 3.20 | 500 | 82.8 | 66.9 | - | 84.8 | 69.0 | - |
6. | BAC-C14 | 4.06 | 50 | 332.0 | 91.1 | 240.3 | - | - | + |
7. | DDAC-C10 | 4.29 | 50 | 326.0 | 186.2 | - | - | - | + |
8. | BAC-C16 | 4.92 | 50 | 360.0 | 91.1 | 268.3 | - | - | + |
9. | Perchlorate | 5.11 | 500 | 98.9 | 82.9 | - | 100.8 | 85.0 | - |
10. | DDAC-C12 | 5.77 | 50 | 382.0 | 214.3 | - | - | - | + |
11. | BAC-C18 | 5.79 | 50 | 388.0 | 91.1 | 296.3 | - | - | + |
Column | Ultra IBD (cat.# 9175312) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 2.1 mm ID | ||||||||||||||||||||
Particle Size: | 3 µm | ||||||||||||||||||||
Pore Size: | 100 Å | ||||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||
Diluent: | Water, 0.5% formic acid | ||||||||||||||||||||
Inj. Vol.: | 1 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | Water, 0.5% formic acid | ||||||||||||||||||||
B: | Acetonitrile, 0.5% formic acid | ||||||||||||||||||||
|
Detector | Shimadzu 8045 MS/MS |
---|---|
Ion Source: | Electrospray |
Ion Mode: | ESI+/ESI- |
Instrument | Shimadzu Nexera X2 |
Sample Preparation | Samples were aliquoted into 2.0 mL, 9 mm screw-thread polypropylene vials (cat.# 23242) and capped with 9 mm solid-top polyethylene caps (cat.# 23244). |