Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

The Resolution of 9-THC, 8-THC, hydroxy- and carboxy- metabolites on Raptor FluoroPhenyl

LC_FF0625
PeakstR (min)PrecursorProduct 1Product 2
1.11-OH-Δ8-THC3.65331.3313.3193.2
2.11-OH-Δ9-THC3.94331.3313.3193.2
3.Δ8-THC-COOH4.07345.0327.2299.2
4.Δ9-THC-COOH4.77345.0327.2299.2
5.Δ8-THC10.81315.0193.2123.2
6.Δ9-THC11.41315.0193.2123.2
ColumnRaptor FluoroPhenyl (cat.# 9319A1E)
Dimensions:100 mm x 3.0 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:30 °C
Standard/SampleΔ9-Tetrahydrocannabinol (Δ9-THC) (cat.# 34067)
(±)11-nor-9-carboxy-Δ-9-THC (Δ9-THC-COOH) (cat.# 34068)
Δ8-Tetrahydrocannabinol (Δ8-THC) (cat.# 34090)
Other compounds obtained separately.
Diluent:37:63 Water:methanol containing 0.1% formic acid
Conc.:100 ng/mL
Inj. Vol.:3 µL
Mobile Phase
A:Water, 0.1% formic acid
B:Methanol, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.93763
12.000.93763
DetectorShimadzu 8060 MS/MS
Ion Source:Electrospray
Ion Mode:ESI+
Mode:MRM
InstrumentShimadzu Nexera X2
Sample PreparationStandards were aliquoted into 2 mL, screw-thread vials (cat.# 21143) and capped with short-cap, screw-vial closures (cat.# 24498).