Peaks | tR (min) | Conc. (ng/mL) | Precursor Ion | Product Ion | Product Ion | |
---|---|---|---|---|---|---|
1. | Lasalocid A | 1.92 | 100 | 613.42 | 377.28 | 595.40 |
2. | Monensin | 2.12 | 100 | 693.50 | 675.44 | 461.30 |
3. | Salinomycin | 2.19 | 100 | 773.57 | 431.24 | 531.39 |
4. | Maduramicin | 2.30 | 100 | 939.65 | 877.58 | 473.31 |
5. | Narasin | 2.58 | 100 | 787.59 | 431.27 | 531.35 |
Column | Raptor Biphenyl (cat.# 9309A12) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 2.1 mm ID | ||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||
Guard Column: | Raptor Biphenyl EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9309A0252) | ||||||||||||||||||||
Temp.: | 40 °C | ||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||
Diluent: | Water:methanol (10:90) | ||||||||||||||||||||
Conc.: | 100 ng/mL | ||||||||||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | 0.5% Formic acid in water | ||||||||||||||||||||
B: | 0.5% Formic acid in acetonitrile | ||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Mode: | ESI+ |
Mode: | MRM |
Instrument | UHPLC |