Peaks | tR (min) | Conc. (ng/mL) | Precursor Ion | Product Ion | Product Ion | |
---|---|---|---|---|---|---|
1. | Oxytetracycline | 1.28 | 25 | 461.27 | 426.15 | 443.32 |
2. | Tetracycline | 1.34 | 25 | 445.28 | 154.07 | 427.32 |
3. | Thiamphenicol* | 1.48 | 200 | 354.16 | 290.04 | 184.98 |
4. | Chlortetracycline | 1.56 | 25 | 479.27 | 154.07 | 371.06 |
5. | Florfenicol* | 1.86 | 200 | 356.10 | 336.02 | 184.98 |
6. | Chloramphenicol* | 1.95 | 200 | 321.16 | 151.99 | 257.04 |
Column | Raptor C18 (cat.# 9304A12) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 2.1 mm ID | ||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||
Guard Column: | Raptor C18 EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9304A0252) | ||||||||||||||||||||
Temp.: | 40 °C | ||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||
Diluent: | Water | ||||||||||||||||||||
Conc.: | 25–200 ng/mL | ||||||||||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | 0.1% Formic acid in water | ||||||||||||||||||||
B: | 0.1% Formic acid in acetonitrile | ||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Mode: | ESI+/ESI- |
Mode: | MRM |
Instrument | UHPLC |
Notes | Tetracyclines and amphenicols were analyzed with ESI+ and ESI- mode, respectively. |