Nitrofurans on Force FluoroPhenyl by UHPLC-MS/MS

LC_FS0507
PeakstR (min)Precursor IonProduct IonProduct Ion
1.Nitrofurantoin0.94239.1121.995.0
2.Nitrofurazone0.99199.1107.954.0
3.Furazolidone1.12226.295.067.0
4.Nifuroxazide1.40276.2121.093.0
ColumnForce FluoroPhenyl (cat.# 9639252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.8 µm
Temp.:40 °C
Standard/Sample
Diluent:Water
Conc.:50 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Acetic acid in water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.000.47030
1.500.4595
1.510.47030
3.500.47030
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC