Peaks | tR (min) | Conc. (ng/mL) | Precursor Ion | Product Ion | Product Ion | |
---|---|---|---|---|---|---|
1. | Bisphenol S | 0.84 | 5.00 | 249.2 | 108.1 | 92.1 |
2. | Bisphenol F | 1.62 | 350 | 199.3 | 93.1 | 105.1 |
3. | Bisphenol E | 2.06 | 100 | 213.3 | 198.3 | 197.4 |
4. | Bisphenol A | 2.50 | 100 | 227.3 | 212.3 | 133.1 |
5. | Bisphenol AF | 2.71 | 2.00 | 335.2 | 265.3 | 177.3 |
6. | Bisphenol B | 3.13 | 100 | 241.3 | 212.4 | 211.3 |
7. | Bisphenol C | 3.43 | 350 | 255.3 | 240.4 | 147.3 |
Peaks | tR (min) | Conc. (ng/mL) | Precursor Ion | Product Ion | Product Ion | |
---|---|---|---|---|---|---|
8. | Bisphenol AP | 3.98 | 25.0 | 289.3 | 274.3 | 273.3 |
9. | Bisphenol Z | 4.25 | 250 | 267.2 | 173.4 | 145.2 |
10. | Bisphenol G | 4.72 | 250 | 311.2 | 295.4 | 296.4 |
11. | Bisphenol FL | 4.90 | 50.0 | 348.8 | 256.2 | - |
12. | Bisphenol BP | 5.14 | 50.0 | 351.2 | 273.3 | 274.3 |
13. | Bisphenol M | 5.39 | 15.0 | 345.2 | 330.3 | 251.4 |
14. | Bisphenol P | 5.67 | 50.0 | 345.2 | 330.4 | 315.3 |
15. | Bisphenol PH | 6.11 | 350 | 379.2 | 209.4 | 364.4 |
Column | Raptor Biphenyl (cat.# 9309252) | ||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 50 mm x 2.1 mm ID | ||||||||||||||||||||
Particle Size: | 1.8 µm | ||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||
Temp.: | 25 °C | ||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||
Diluent: | 75:25 Water:methanol | ||||||||||||||||||||
Conc.: | 2.00-350 ng/mL | ||||||||||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||
A: | Water | ||||||||||||||||||||
B: | Methanol | ||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Mode: | ESI- |
Mode: | MRM |
Instrument | UHPLC |