Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | tR (min) | |
---|---|---|
1. | Cannabigerolic acid (CBGA) | 1.19 |
2. | Δ9-Tetrahydrocannabinol (Δ9-THC) | 2.49 |
3. | Tetrahydrocannabinolic acid A (THCA-A) | 3.27 |
Column | Raptor ARC-18 (cat.# 931421E) | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 3.0 mm ID | ||||||||||||
Particle Size: | 1.8 µm | ||||||||||||
Pore Size: | 90 Å | ||||||||||||
Guard Column: | UltraShield UHPLC precolumn filter (cat.# 25809) | ||||||||||||
Temp.: | 30 °C | ||||||||||||
Standard/Sample | |||||||||||||
Diluent: | 25:75 Water:methanol | ||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||
Mobile Phase | |||||||||||||
A: | Water, 5 mM ammonium formate, 0.1% formic acid | ||||||||||||
B: | Acetonitrile, 0.1% formic acid | ||||||||||||
|
Detector | UV/Vis @ 228 nm |
---|---|
Instrument | UHPLC |
Sample Preparation | 200 mg of ground flower was weighed into a 50 mL centrifuge tube. 10 mL of IPA was added prior to vortexing (5 seconds) and sonicating (5 minutes) over a total of three cycles. The sample was then centrifuged for 5 minutes at 3000 rpm. The supernatant was diluted 50-fold in 25:75 water:methanol, vortexed briefly, and filtered using a 0.2 μm Thomson SINGLE StEP standard filter vial (cat.#: 25893) prior to analysis. |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.