Peaks | tR (min) | |
---|---|---|
1. | Cannabidivarinic acid (CBDVA) | 1.570 |
2. | Cannabidivarin (CBDV) | 1.728 |
3. | Cannabidiolic acid (CBDA) | 2.159 |
4. | Cannabigerolic acid (CBGA) | 2.304 |
5. | Cannabigerol (CBG) | 2.411 |
6. | Cannabidiol (CBD) | 2.537 |
7. | Tetrahydrocannabivarin (THCV) | 2.770 |
8. | Cannabichromevarin (CBCV) | 3.355 |
9. | Tetrahydrocannabivarinic acid (THCVA) | 3.507 |
Peaks | tR (min) | |
---|---|---|
10. | Cannabinol (CBN) | 3.756 |
11. | Cannabinolic acid (CBNA) | 4.471 |
12. | Δ9-Tetrahydrocannabinol (Δ9-THC) | 4.706 |
13. | Δ8-Tetrahydrocannabinol (Δ8-THC) | 4.865 |
14. | Cannabicyclol (CBL) | 5.594 |
15. | Cannabichromene (CBC) | 5.878 |
16. | Tetrahydrocannabinolic acid A (THCA-A) | 6.213 |
17. | Cannabichromenic acid (CBCA) | 6.965 |
18. | Cannabicyclolic acid (CBLA) | 7.370 |
19. | Cannabicitran (CBT) | 9.359 |
Column | Raptor ARC-18 (cat.# 9314A62) | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 150 mm x 2.1 mm ID | ||||||||||||
Particle Size: | 2.7 µm | ||||||||||||
Pore Size: | 90 Å | ||||||||||||
Guard Column: | Raptor ARC-18 EXP guard column cartridge 2.7 µm (cat.# 9314A0252) | ||||||||||||
Temp.: | 30 °C | ||||||||||||
Standard/Sample | Cannabidivarin (CBDV) (cat.# 34123) | ||||||||||||
Cannabidiolic acid (CBDA) (cat.# 34094) | |||||||||||||
Cannabigerolic acid (CBGA) (cat.# 34112) | |||||||||||||
Cannabigerol (CBG) (cat.# 34091) | |||||||||||||
Cannabidiol (CBD) (cat.# 34011) | |||||||||||||
Tetrahydrocannabivarin (THCV) (cat.# 34100) | |||||||||||||
Cannabinol (CBN) (cat.# 34010) | |||||||||||||
d9-Tetrahydrocannabinol (d9-THC) (cat.# 34067) | |||||||||||||
d8-Tetrahydrocannabinol (d8-THC) (cat.# 34090) | |||||||||||||
Cannabichromene (CBC) (cat.# 34092) | |||||||||||||
d9-Tetrahydrocannabinolic acid A (THCA-A) (cat.# 34093) | |||||||||||||
Cannabicyclol (CBL) (cat.# 34130) | |||||||||||||
Compounds not present in these mixes were obtain separately. | |||||||||||||
Diluent: | Methanol | ||||||||||||
Conc.: | 50 ppm | ||||||||||||
Inj. Vol.: | 2 µL | ||||||||||||
Mobile Phase | |||||||||||||
A: | Water, 5 mM ammonium formate, 0.1% formic acid | ||||||||||||
B: | Acetonitrile, 0.1% formic acid | ||||||||||||
|
Detector | UV/Vis @ 228 nm |
---|---|
Instrument | UHPLC |