Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | tR (min) | |
---|---|---|
1. | Cannabidivarinic acid (CBDVA) | 1.775 |
2. | Cannabidivarin (CBDV) | 1.973 |
3. | Cannabidiolic acid (CBDA) | 2.442 |
4. | Cannabigerolic acid (CBGA) | 2.575 |
5. | Cannabigerol (CBG) | 2.732 |
6. | Cannabidiol (CBD) | 2.904 |
7. | Tetrahydrocannabivarin (THCV) | 3.203 |
8. | Cannabichromevarin (CBCV) | 3.867 |
9. | Tetrahydrocannabivarinic acid (THCVA) | 4.023 |
10. | Cannabinol (CBN) | 4.321 |
Peaks | tR (min) | |
---|---|---|
11. | Cannabidiphorol (CBDP) | 4.866 |
12. | Cannabinolic acid (CBNA) | 5.099 |
13. | Δ9-Tetrahydrocannabinol (Δ9-THC) | 5.460 |
14. | Δ8-Tetrahydrocannabinol (Δ8-THC) | 5.637 |
15. | Cannabicyclol (CBL) | 6.496 |
16. | Cannabichromene (CBC) | 6.788 |
17. | Tetrahydrocannabinolic acid A (THCA-A) | 7.122 |
18. | Cannabichromenic acid (CBCA) | 7.926 |
19. | Cannabicyclolic acid (CBLA) | 8.458 |
20. | Tetrahydrocannabiphorol (THCP) | 10.251 |
21. | Cannabicitran (CBT) | 11.015 |
Column | Raptor ARC-18 (cat.# 9314A65) | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 150 mm x 4.6 mm ID | ||||||||||||
Particle Size: | 2.7 µm | ||||||||||||
Pore Size: | 90 Å | ||||||||||||
Guard Column: | Raptor ARC-18 5 mm, 4.6 mm ID, 2.7 µm (cat.# 9314A0250) | ||||||||||||
Temp.: | 30 °C | ||||||||||||
Standard/Sample | Cannabidivarinic acid (CBDVA) (cat.# 34134) | ||||||||||||
Cannabidivarin (CBDV) (cat.# 34123) | |||||||||||||
Cannabidiolic acid (CBDA) (cat.# 34094) | |||||||||||||
Cannabigerolic acid (CBGA) (cat.# 34135) | |||||||||||||
Cannabigerol (CBG) (cat.# 34091) | |||||||||||||
Cannabidiol (CBD) (cat.# 34011) | |||||||||||||
Tetrahydrocannabivarin (THCV) (cat.# 34100) | |||||||||||||
Cannabinol (CBN) (cat.# 34010) | |||||||||||||
d9-Tetrahydrocannabinol (d9-THC) (cat.# 34067) | |||||||||||||
d8-Tetrahydrocannabinol (d8-THC) (cat.# 34090) | |||||||||||||
Cannabicyclol (CBL) (cat.# 34130) | |||||||||||||
Cannabichromene (CBC) (cat.# 34092) | |||||||||||||
d9-Tetrahydrocannabinolic acid A (THCA-A) (cat.# 34111) | |||||||||||||
Compounds not present in these mixes were obtained separately. | |||||||||||||
Diluent: | Methanol | ||||||||||||
Conc.: | 50 ppm | ||||||||||||
Inj. Vol.: | 5 µL | ||||||||||||
Mobile Phase | |||||||||||||
A: | Water, 5 mM ammonium formate, 0.1% formic acid | ||||||||||||
B: | Acetonitrile, 0.1% formic acid | ||||||||||||
|
Detector | UV/Vis @ 228 nm |
---|---|
Instrument | UHPLC |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.