Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.
Peaks | tR (min) | |
---|---|---|
1. | Cannabidivarinic acid (CBDVA) | 0.813 |
2. | Cannabidivarin (CBDV) | 0.888 |
3. | Cannabidiolic acid (CBDA) | 1.102 |
4. | Cannabigerolic acid (CBGA) | 1.172 |
5. | Cannabigerol (CBG) | 1.222 |
6. | Cannabidiol (CBD) | 1.278 |
7. | Tetrahydrocannabivarin (THCV) | 1.398 |
8. | Cannabichromevarin (CBCV) | 1.679 |
9. | Tetrahydrocannabivarinic acid (THCVA) | 1.768 |
10. | Cannabinol (CBN) | 1.872 |
Peaks | tR (min) | |
---|---|---|
11. | Cannabidiphorol (CBDP) | 2.088 |
12. | Cannabinolic acid (CBNA) | 2.222 |
13. | Δ9-Tetrahydrocannabinol (Δ9-THC) | 2.325 |
14. | Δ8-Tetrahydrocannabinol (Δ8-THC) | 2.394 |
15. | Cannabicyclol (CBL) | 2.740 |
16. | Cannabichromene (CBC) | 2.882 |
17. | Tetrahydrocannabinolic acid A (THCA-A) | 3.065 |
18. | Cannabichromenic acid (CBCA) | 3.418 |
19. | Cannabicyclolic acid (CBLA) | 3.620 |
20. | Tetrahydrocannabiphorol (THCP) | 4.273 |
21. | Cannabicitran (CBT) | 4.552 |
Column | Raptor ARC-18 (cat.# 931421E) | ||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 3.0 mm ID | ||||||||||||
Particle Size: | 1.8 µm | ||||||||||||
Pore Size: | 90 Å | ||||||||||||
Guard Column: | UltraShield UHPLC precolumn filter, 0.2 μm frit (cat.# 25809) | ||||||||||||
Temp.: | 30 °C | ||||||||||||
Standard/Sample | Cannabidivarinic acid (CBDVA) (cat.# 34134) | ||||||||||||
Cannabidivarin (CBDV) (cat.# 34123) | |||||||||||||
Cannabidiolic acid (CBDA) (cat.# 34094) | |||||||||||||
Cannabigerolic acid (CBGA) (cat.# 34135) | |||||||||||||
Cannabigerol (CBG) (cat.# 34091) | |||||||||||||
Cannabidiol (CBD) (cat.# 34011) | |||||||||||||
Tetrahydrocannabivarin (THCV) (cat.# 34100) | |||||||||||||
Cannabinol (CBN) (cat.# 34010) | |||||||||||||
d9-Tetrahydrocannabinol (d9-THC) (cat.# 34067) | |||||||||||||
d8-Tetrahydrocannabinol (d8-THC) (cat.# 34090) | |||||||||||||
Cannabicyclol (CBL) (cat.# 34130) | |||||||||||||
Cannabichromene (CBC) (cat.# 34092) | |||||||||||||
d9-Tetrahydrocannabinolic acid A (THCA-A) (cat.# 34111) | |||||||||||||
Compounds not present in these mixes were obtained separately. | |||||||||||||
Diluent: | Methanol | ||||||||||||
Conc.: | 50 ppm | ||||||||||||
Inj. Vol.: | 1 µL | ||||||||||||
Mobile Phase | |||||||||||||
A: | Water, 5 mM ammonium formate, 0.1% formic acid | ||||||||||||
B: | Acetonitrile, 0.1% formic acid | ||||||||||||
|
Detector | UV/Vis @ 228 nm |
---|---|
Instrument | UHPLC |
We have determined your location as:
Or select your location below:
Sorry! We are unable to determine your location. Please select your location below.