Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Cannabinoids Acids 7 Standard on Raptor ARC-18 2.7 μm by LC-UV

LC_GN0679
PeakstR (min)Conc.
(µg/mL)
1.Cannabidivarinic acid (CBDVA)1.86050
2.Cannabidiolic acid (CBDA)2.61050
3.Cannabigerolic acid (CBGA)2.78550
4.Tetrahydrocannabivarinic acid (THCVA)4.32550
5.Cannabinolic acid (CBNA)5.55050
6.Tetrahydrocannabinolic acid A (THCA-A)7.81050
7.Cannabichromenic acid (CBCA)8.78050
ColumnRaptor ARC-18 (cat.# 9314A65)
Dimensions:150 mm x 4.6 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor ARC-18 5 mm, 4.6 mm ID, 2.7 µm (cat.# 9314A0250)
Temp.:30 °C
Standard/SampleCannabinoids acids 7 standard (cat.# 34144)
Diluent:Acetonitrile
Conc.:50 ppm
Inj. Vol.:5 µL
Mobile Phase
A:Water, 5 mM ammonium formate, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.001.52575
10.001.52575
DetectorUV/Vis @ 228 nm
InstrumentUHPLC
Sample PreparationStandards were aliquoted into 2 mL, screw-thread vials (cat.# 21143) and capped with short-cap, screw-vial closures (cat.# 24498).