Peaks | tR (min) | Precursor Ion | Product Ion 1 | Product Ion 2 | Polarity | |
---|---|---|---|---|---|---|
1. | Daminozide | 1.2 | 161.1 | 143.2 | 44.1 | + |
2. | Acephate | 2.5 | 184.0 | 143.1 | 95.1 | + |
3. | Dinotefuran | 2.7 | 203.2 | 129.3 | 73.1 | + |
4. | Oxamyl | 2.8 | 237.1 | 72.1 | 90.1 | + |
5. | Flonicamid | 2.9 | 230.1 | 203.1 | 174.1 | + |
6. | Methomyl | 2.9 | 163.1 | 88.1 | 106.1 | + |
7. | Thiamethoxam | 2.9 | 292.0 | 211.1 | 181.1 | + |
8. | Imidacloprid | 3.1 | 256.1 | 209.1 | 175.1 | + |
9. | Clothianidin | 3.2 | 250.1 | 169.2 | 132.3 | + |
10. | Mevinphos I | 3.3 | 225.1 | 127.1 | 193.2 | + |
11. | Dimethoate | 3.3 | 230.0 | 199.1 | 125.1 | + |
12. | Acetamiprid | 3.4 | 223.0 | 126.1 | 56.1 | + |
13. | Thiacloprid | 3.6 | 253.0 | 126.0 | 90.1 | + |
14. | Mevinphos II | 3.7 | 225.1 | 127.1 | 193.2 | + |
15. | Aldicarb | 4.0 | 208.4 | 116.0 | 89.0 | + |
16. | Pirimicarb | 4.1 | 239.2 | 72.1 | 182.3 | + |
17. | Thiophanate-methyl | 4.5 | 343.1 | 151.3 | 311.2 | + |
18. | Dichlorvos | 4.6 | 220.9 | 108.8 | 79.2 | + |
19. | Propoxur | 4.7 | 210.1 | 111.1 | 93.1 | + |
20. | Carbofuran | 4.7 | 222.1 | 123.1 | 165.2 | + |
21. | Carbaryl | 5.0 | 202.1 | 145.1 | 127.1 | + |
22. | Imazalil | 5.1 | 297.0 | 159.0 | 201.0 | + |
23. | Cyantraniliprole | 5.2 | 475.1 | 285.9 | 444.2 | + |
24. | Azadirachtin | 5.4 | 703.4 | 567.3 | 585.3 | + |
25. | Dodemorph | 5.7 | 282.3 | 116.2 | 98.1 | + |
26. | Metalaxyl | 5.9 | 280.2 | 220.2 | 192.2 | + |
27. | Naled | 5.9 | 397.8 | 127.1 | 109.1 | + |
28. | Chlorantraniliprole | 6.1 | 483.9 | 452.9 | 285.9 | + |
29. | Phosmet | 6.1 | 318.0 | 160.1 | 77.2 | + |
30. | Spiroxamine | 6.3 | 298.3 | 144.2 | 100.2 | + |
31. | Azoxystrobin | 6.5 | 404.0 | 372.1 | 344.1 | + |
32. | Methiocarb | 6.5 | 226.1 | 169.1 | 121.1 | + |
33. | Fludioxonil | 6.6 | 274.0 | 180.0 | 126.0 | - |
34. | Dimethomorph I | 6.7 | 388.2 | 301.2 | 165.3 | + |
35. | Boscalid | 6.7 | 342.9 | 307.5 | 140.1 | + |
36. | Paclobutrazol | 6.9 | 294.3 | 70.1 | 125.1 | + |
37. | Malathion | 6.9 | 331.0 | 127.2 | 285.2 | + |
38. | Dimethomorph II | 7.1 | 388.2 | 301.2 | 165.3 | + |
39. | Myclobutanil | 7.2 | 289.1 | 70.1 | 125.1 | + |
40. | Bifenazate | 7.4 | 301.0 | 198.1 | 170.2 | + |
41. | Fluopyram | 7.5 | 397.1 | 173.2 | 208.3 | + |
42. | Cyprodinil | 7.7 | 226.2 | 93.1 | 108.2 | + |
43. | Spirotetramat | 8.0 | 374.2 | 216.1 | 302.1 | + |
44. | Ethoprophos | 8.2 | 243.1 | 131.1 | 97.1 | + |
45. | Iprodione | 8.4 | 330.1 | 245.0 | + | |
46. | Iprodione (qualifier) | 8.4 | 332.2 | 247.0 | + | |
47. | Fipronil | 8.5 | 436.8 | 331.8 | 251.9 | - |
48. | Fenoxycarb | 8.8 | 302.1 | 88.1 | 116.1 | + |
Peaks | tR (min) | Precursor Ion | Product Ion 1 | Product Ion 2 | Polarity | |
---|---|---|---|---|---|---|
49. | Tetrachlorvinphos | 9.0 | 365.1 | 127.2 | 204.0 | + |
50. | Tebufenozide | 9.0 | 353.2 | 133.2 | 297.2 | + |
51. | Kresoxim-methyl | 9.1 | 314.2 | 267.2 | 222.2 | + |
52. | Fenthion | 9.3 | 279.0 | 247.3 | 138.4 | + |
53. | Tebuconazole | 9.4 | 308.1 | 70.1 | 125.1 | + |
54. | Benzovindiflupyr | 9.5 | 398.2 | 378.0 | 322.1 | + |
55. | Diazinon | 9.7 | 305.1 | 169.2 | 153.2 | + |
56. | Coumaphos | 10.0 | 363.1 | 227.1 | 307.1 | + |
57. | Propiconazole | 10.0 | 342.0 | 159.0 | 69.2 | + |
58. | Clofentezine | 10.1 | 303.0 | 138.1 | 102.1 | + |
59. | Pyraclostrobin | 10.3 | 388.2 | 194.4 | 163.4 | + |
60. | Spinosad (Spinosyn A) | 10.6 | 732.4 | 142.2 | 98.1 | + |
61. | MGK-264 I | 11.1 | 276.2 | 210.1 | 121.1 | + |
62. | Prallethrin | 11.5 | 301.2 | 123.2 | 104.9 | + |
63. | Trifloxystrobin | 11.6 | 409.2 | 186.1 | 145.1 | + |
64. | Spinosad (Spinosyn D) | 11.7 | 746.5 | 142.3 | 98.4 | + |
65. | Spinetoram (Spinosyn J) | 11.9 | 748.5 | 142.5 | 98.3 | + |
66. | Novaluron | 12.0 | 493.2 | 158.1 | 140.9 | + |
67. | Pyrethrins (Pyrethrin II) | 12.0 | 373.1 | 161.1 | 133.2 | + |
68. | MGK-264 II | 12.0 | 276.2 | 210.1 | 121.1 | + |
69. | Buprofezin | 12.3 | 306.2 | 201.4 | 57.1 | + |
70. | Teflubenzuron | 12.4 | 379.0 | 338.9 | 196.1 | - |
71. | Spinetoram (Spinosyn L) | 13.0 | 760.5 | 142.2 | 98.1 | + |
72. | Piperonyl butoxide | 13.2 | 356.3 | 177.2 | 119.2 | + |
73. | Chlorpyrifos | 13.4 | 349.9 | 198.0 | 97.1 | + |
74. | Allethrin | 13.4 | 303.3 | 135.3 | 123.2 | + |
75. | Hexythiazox | 13.6 | 353.1 | 228.1 | 168.1 | + |
76. | Etoxazole | 14.5 | 360.2 | 141.1 | 304.2 | + |
77. | Spiromesifen | 14.6 | 388.4 | 273.3 | 255.3 | + |
78. | Pyrethrins (Pyrethrin I) | 15.0 | 329.2 | 161.2 | 133.2 | + |
79. | Cyfluthrin | 15.2 | 451.1 | 191.2 | + | |
80. | Cyfluthrin (qualifier) | 15.2 | 453.1 | 193.2 | + | |
81. | Spirodiclofen | 15.5 | 411.2 | 71.2 | 43.4 | + |
82. | Cypermethrin | 15.6 | 433.1 | 191.0 | 416.0 | + |
83. | Fenpyroximate | 15.6 | 422.2 | 366.1 | 138.1 | + |
84. | Deltamethrin | 15.9 | 523.1 | 280.8 | 505.9 | + |
85. | Fenvalerate | 16.1 | 436.9 | 167.0 | + | |
86. | Pyridaben | 16.2 | 365.1 | 147.2 | 309.2 | + |
87. | Resmethrin | 16.4 | 339.3 | 171.1 | 128.1 | + |
88. | Permethrin trans | 16.5 | 408.3 | 183.2 | 355.1 | + |
89. | Methoprene | 16.7 | 311.3 | 191.3 | 279.1 | + |
90. | Phenothrin (trans) | 16.7 | 351.3 | 249.2 | 182.9 | + |
91. | Permethrin-cis | 16.9 | 408.3 | 183.2 | 355.1 | + |
92. | Abamectin | 16.9 | 890.5 | 305.4 | 567.4 | + |
93. | d-Phenothrin (cis) | 16.9 | 351.3 | 249.2 | 182.9 | + |
94. | Etofenprox | 17.0 | 394.3 | 177.2 | 359.3 | + |
95. | Bifenthrin | 17.1 | 440.0 | 181.2 | 166.2 | + |
96. | Acequinocyl | 18.2 | 402.3 | 343.2 | 189.0 | + |
Column | Raptor ARC-18 (cat.# 9314A62) | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 150 mm x 2.1 mm ID | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Pore Size: | 90 Å | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Guard Column: | Raptor ARC-18 EXP guard column cartridge 5 mm, 2.1 mm ID, 2.7 µm (cat.# 9314A0252) | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Temp.: | 40 °C | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Standard/Sample | Canadian pesticide standard #1 (cat.# 32593) | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Canadian pesticide standard #2 (cat.# 32594) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Canadian pesticide standard #3 (cat.# 32595) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Canadian pesticide standard #4 (cat.# 32596) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Canadian pesticide standard #5 (cat.# 32597) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Canadian pesticide standard #6 (cat.# 32598) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Azadirachtin (cat.# 32599) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Dichlorvos (DDVP) (cat.# 32585) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Diluent: | Acetonitrile | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Conc.: | 100 ng/mL | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Inj. Vol.: | 1 µL | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
A: | Water, 2 mM ammonium formate, 0.1% formic acid | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
B: | Methanol, 2 mM ammonium formate, 0.1% formic acid | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Detector | MS/MS |
---|---|
Ion Source: | Electrospray |
Ion Mode: | ESI+/ESI- |
Mode: | MRM |
Instrument | Shimadzu 8045 |
Sample Preparation | Standards were aliquoted into 2 mL, screw-thread vials (cat.# 21143) and capped with short-cap, screw-vial closures (cat.# 24498). |
Notes | Autosampler temperature: 10 °C Canadian pesticide standards are also available as a kit: Canadian pesticide kit (cat.# 32592) |