Peaks | tR (min) | |
---|---|---|
1. | N,N-desmethyl Psilocybin (Norbaeocystin) | 1.691 |
2. | N-desmethyl Psilocybin (Baeocystin) | 2.424 |
3. | Psilocybin | 3.346 |
4. | 4-hydroxy-N-Methyltryptamine (Norpsilocin) | 3.917 |
5. | 4-hydroxy-N,N-Dimethyltryptamine (Psilocin) | 4.575 |
Column | Force Biphenyl (cat.# 962931E) | ||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Dimensions: | 100 mm x 3.0 mm ID | ||||||||||||||||||||||||||||
Particle Size: | 3.0 µm | ||||||||||||||||||||||||||||
Pore Size: | 100 Å | ||||||||||||||||||||||||||||
Guard Column: | Force Biphenyl EXP Guard Cartridge 5.0 mm, 3.0 mm ID (cat.# 962950253) | ||||||||||||||||||||||||||||
Temp.: | 50 °C | ||||||||||||||||||||||||||||
Standard/Sample | |||||||||||||||||||||||||||||
Diluent: | Water | ||||||||||||||||||||||||||||
Inj. Vol.: | 3 µL | ||||||||||||||||||||||||||||
Mobile Phase | |||||||||||||||||||||||||||||
A: | Water, 10 mM ammonium formate, 0.1% formic acid | ||||||||||||||||||||||||||||
B: | Methanol, 0.1% formic acid | ||||||||||||||||||||||||||||
|
Detector | UV/Vis @ 222 nm |
---|---|
Instrument | Agilent 1100 |
Sample Preparation | 500 mg of dried, homogenized mushroom was weighed into a 50-mL Environmental Express tube (6.0 µm). 20 mL of water with 1% acetic acid (v/v) was added, tube capped, and vortexed for 5 min at 2500 RPM. The sample was then filtered with a plunger. An aliquot was obtained using a 3-mL syringe (cat.#22773) equipped with a 0.22 µm PTFE syringe filter (cat.# 23984). The sample was then diluted 5x with HPLC grade water into a 2.0 mL amber, short-cap vial (cat.# 21142) and capped with a 9 mm short cap (cat.#24497). |
Notes | The guard cartridge was equipped with an UltraShield precolumn filter, 0.22 µm (cat.# 25809). System and column were passivated using LC Passivation Solution (cat.# 32475). The outlined method conditions were used to perform five injections of passivation solution (5 µL) on column, diverting to waste. Five warm-up injections were completed prior to analysis. Storage Conditions: Flush column using acetonitrile with 0.1% formic acid for 5 min. |
Acknowledgement | We would like to thank 2 River Labs for their use and accommodations of lab space, instruments, and samples during collaboration. |