Chlormequat, Mepiquat, Paraquat, and Diquat on Raptor HILIC-Si by LC-MS/MS

LC_GN0696
PeakstR (min)Precursor Ion 1Product Ion 1 Precursor Ion 2Product Ion 2
1.Chlormequat1.325122.058.2122.063.1
2.Mepiquat Chloride1.724114.098.2114.058.2
3.Paraquat6.036186.0171.0171.077.2
4.Diquat6.491183.0157.0183.0130.0
ColumnRaptor HILIC-Si (cat.# 9310A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:60 °C
Standard/SampleParaquat & Diquat Calibration Mix (cat.# 32437)
Compounds not present in these mixes were obtained separately.
Diluent:Acetonitrile:water 90:10, 5 mM ammonium formate, 0.5% formic acid
Conc.:100 ng/mL
Inj. Vol.:1 µL
Mobile Phase
A:Water, 50 mM ammonium formate, 0.5% formic acid
B:Acetonitrile:water 90:10, 5 mM ammonium formate, 0.5% formic acid
Time (min)Flow (mL/min)%A%B
0.000.60100
4.000.62575
7.000.62575
7.010.60100
10.000.60100
DetectorShimadzu 8060 MS/MS
Ion Source:Electrospray
Ion Mode:ESI+
Mode:MRM
InstrumentShimadzu Nexera X2
Sample PreparationSamples were aliquoted into a 2.0 mL, 9 mm Screw-Thread Polypropylene Vials (cat.# 23242) and capped with 9 mm Solid-Top Polyethylene Caps (cat.# 23244).