Specifications for Super Clean Gas Filters
Restek Super Clean Traps measure: 10 5/8" x 1 3/4" (27 x 4.4 cm).
-----Capacity------ | |||||||||
Type of Filter | Outlet Gas Quality (%) | Maximum Pressure/ Maximum Flow Rates |
Use for: | Indicator Color Change | H20 (g) | O2 (mL) | Hydrocarbons (g) (as n-butane) |
Carbon Dioxide (g) | Estimated Lifetime (years) |
Moisture cat.# 22028 |
>99.9999 | 15 bar 217 psi/ 7 L/min. |
Inert carrier gas Air Hydrogen |
Yellow/orange to clear | 7.2 | — | — | — | >2 |
Oxygen cat.# 22029 |
>99.9999 | 15 bar 217 psi/ 7 L/min. |
Inert carrier gas Hydrogen |
Green to grey | NA | 150 | — | — | >2 |
Hydrocarbons cat.# 22030 |
>99.9999 | 15 bar 217 psi/ 7 L/min. |
Inert carrier gas Air Hydrogen |
No indicator | NA | — | 12 | — | >2 |
Fuel Gas1 cat.# 22022 |
>99.9999 | 15 bar 217 psi/ 7 L/min. |
Inert carrier gas Air Hydrogen |
Yellow/orange to clear | 3.5 | — | 6 | — | >1.5 |
Triple2 cat.# 22020 |
>99.9999 | 15 bar 217 psi/ 7 L/min. |
Inert carrier gas Hydrogen |
Yellow/orange to clear Green to grey |
1.8 | 75 | 4 | — | >1 |
Helium Specific2 cat.# 21982 |
>99.9999 | 15 bar 217 psi/ 7 L/min. |
Helium | Yellow/orange to clear Green to grey |
1.8 | 75 | 4 | — | >1 |
CO2, Sulfur, Moisture cat.# 23843 |
>99.9999 | 15 bar 217 psi/ 7 L/min. |
Inert carrier gas Air Hydrogen |
Yellow/orange to clear Grey to white |
3.5 | — | — | 6 | >2 |
1 Removes hydrocarbons, moisture.
2 Removes hydrocarbons, moisture, oxygen.
Note:
Restek replacement filters fit all cartridge filter baseplates regardless of manufacturer. However, if the baseplate was manufactured by Agilent or Varian, the retaining nut that came with the original filter will not work and will require a universal ring nut. Please contact customer service to receive a complimentary universal ring nut.
Each baseplate unit measures: 4" x 4" x 1 7/8" (10.2 x 10.2 x 4.8 cm).
Standard baseplate inlet/outlet fittings accept 1/8" tubing. To adapt to 1/4" tubing, order 1/8" to 1/4" tube end reducers.