Featured Application: 19 Cannabinoids on Raptor ARC-18

Fast, Low-Solvent Analysis of Cannabinoids Increases Lab Productivity and Decreases Solvent Costs

  • Leaner, greener method uses only 3 mL of acetonitrile per run.
  • 10-min analysis of 19 cannabinoids increases sample throughput and profiling capability.
  • Simple, isocratic method is more transferable between systems and labs than gradient methods.
 

In the highly competitive landscape of cannabis testing, labs that can expand services while decreasing costs have a strong advantage. Potency testing and cannabinoid profiling are high-demand tests that can be improved by increasing speed, expanding analyte lists, and reducing costs. The low-solvent analysis of cannabinoids shown here accomplishes this by using a Raptor ARC-18 column to fully resolve 19 cannabinoids in a fast 10-minute analysis that uses only 3 mL of acetonitrile per run. The wide range of cannabinoids analyzed ensures more accurate potency results and provides more comprehensive data for product branding and identification. Reducing expensive mobile phase organic solvent consumption significantly lowers the cost per sample and also reduces solvent waste and its associated disposal costs. In addition, this low-solvent analysis of cannabinoids uses a simple isocratic mobile phase, so it is more easily transferrable between instruments compared to more complex methods that incorporate mobile phase gradients or additives.

ColumnRaptor ARC-18 (cat.# 9314A62)
Dimensions:150 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor ARC-18 EXP guard column cartridge 2.7 µm (cat.# 9314A0252)
Temp.:30 °C
Standard/SampleCannabidivarin (CBDV) (cat.# 34123)
Cannabidiolic acid (CBDA) (cat.# 34094)
Cannabigerolic acid (CBGA) (cat.# 34112)
Cannabigerol (CBG) (cat.# 34091)
Cannabidiol (CBD) (cat.# 34011)
Tetrahydrocannabivarin (THCV) (cat.# 34100)
Cannabinol (CBN) (cat.# 34010)
d9-Tetrahydrocannabinol (d9-THC) (cat.# 34067)
d8-Tetrahydrocannabinol (d8-THC) (cat.# 34090)
Cannabichromene (CBC) (cat.# 34092)
d9-Tetrahydrocannabinolic acid A (THCA-A) (cat.# 34093)
Cannabicyclol (CBL) (cat.# 34130)
Compounds not present in these mixes were obtain separately.
Diluent:Methanol
Conc.:50 ppm
Inj. Vol.:2 µL
Mobile Phase
A:Water, 5 mM ammonium formate, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.000.42575
10.000.42575
DetectorUV/Vis @ 228 nm
InstrumentUHPLC
FFFA3123-UNV

Related Products

Raptor ARC-18 LC Columns (USP L1)

View Product

Raptor EXP Guard Column Cartridges

View Product

EXP Direct Connect Holder

View Product

Thomson SINGLE StEP Standard Filter Vials

View Product

Bluestem Glass Solvent Filter

View Product

Tetrahydrocannabivarin (THCV) Standard

View Product

Cannabidiolic Acid (CBDA) Standard

View Product

Cannabichromene (CBC) Standard

View Product

Cannabigerol (CBG) Standard

View Product

d8-Tetrahydrocannabinol (d8-THC) Standard

View Product

d9-Tetrahydrocannabinol (d9-THC) Standard

View Product

Cannabinol (CBN) Standard

View Product

Cannabidiol (CBD) Standard

View Product

Cannabidivarin (CBDV) Standard

View Product

Cannabicyclol (CBL) Standard

View Product