Syringe Filters Solvent Compatibility Chart
- Share:
- Download PDF
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF |
ACIDS | |||||
Acetic, 5% | L | R | R | R | R |
Acetic, 10% | L | R | R | R | R |
Acetic, 25% | N | L | R | R | R |
Acetic, Glacial | N | N | R | R | R |
Boric | – | L | – | R | – |
Formic 25% | L | N | – | R | – |
Hydrochloric 15% | L | L | R | R | L |
Hydrochloric 25% | N | N | R | R | – |
Hydrochloric concentrated | N | N | L | R | N |
Hydrofluoric 10% | N | N | – | – | – |
Hydrofluoric 35% | N | N | – | R | – |
Nitric 25% | N | N | R | R | – |
Nitric 6N, 38% | N | N | L | R | R |
Nitric concentrated | N | N | N | R | N |
Phosphoric 25% | L | N | R | R | – |
Sulfuric 25% | N | N | N | R | – |
Sulfuric 6N, 29% | N | N | N | R | – |
Sulfuric concentrated | N | N | N | R | N |
Trichloroacetic 10% | N | N | – | R | R |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF |
ALKALINES | |||||
Ammonium Hydroxide 25% | N | R | R | R | L |
Formalin 30% | L | L | R | – | – |
Sodium Hydroxide 3N, 12% | N | R | R | R | R |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF |
ALCOHOLS | |||||
Amyl Alcohol | L | R | N | R | R |
Benzyl Alcohol | L | L | L | L | L |
Butyl Alcohol | L | R | L | R | R |
Butyl Cellosolve | N | L | – | L | – |
Ethanol 70% | L | R | L | R | R |
Ethanol 98% | N | R | N | R | R |
Ethylene glycol | L | R | R | R | R |
Glycerol | L | R | R | R | R |
Isobutyl Alcohol | L | L | L | L | L |
Isopropanol, n-Propanol | L | R | R | R | R |
Methanol 98% | N | R | L | R | R |
Methyl Cellosolve | L | L | – | L | – |
Propylene glycol | L | – | R | R | R |
Phenol, Aqueous 10% | – | R | – | R | R |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF |
HYDROCARBONS | |||||
Hexane | L | R | L | R | R |
Xylene | L | R | N | R | R |
Kerosene, Gasoline | L | R | R | R | R |
Tetrakin, Decalin | N | R | – | R | R |
Toluene, benzene | L | R | N | R | R |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF |
HALOGENATED HYDROCARBONS | |||||
Carbon Tetrachloride | N | N | N | N | N |
Chloroform | N | N | N | R | R |
Methylene Chloride | N | L | N | R | N |
Monochlorobenzene | N | N | – | R | – |
Trichloroethylene | N | N | N | R | R |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF |
KETONES | |||||
Acetone | N | R | N | R | N |
Cyclohexanone | N | L | N | R | N |
Isopropylacetone | – | R | – | R | N |
Methyl Ethyl Ketone | N | R | N | R | N |
Methyl Isobutyl Ketone | N | R | – | R | N |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF |
ESTERS | |||||
Amyl Acetate | N | R | L | R | – |
Amyl Propyl & Butyl Acetate | L | – | – | R | – |
Benzyl Benzoate | – | – | – | R | – |
Butyl Acetate | N | – | N | – | – |
Ethyl Acetate & Methyl Acetate | N | R | N | R | R/L |
Isopropyl Myristate | – | – | – | R | – |
Methyl Cellosolve Acetate | N | – | – | R | – |
Propylene Glycol Acetate | – | – | – | R | – |
Tricresyl Phosphate | – | – | – | R | – |
Isopropyl Acetate | L | R | – | – | R |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF |
OXIDES - ETHERS | |||||
Dimethylsulfoxide (DMSO) | N | R | N | R | N |
Dioxane & Tetrahydrofuran | N | R | L | R | L |
Ethyl Ether | L | R | R | R | R |
Isopropyl Ether | – | – | – | R | R |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF |
SOLVENTS WITH NITROGEN | |||||
Acetonitrile | N | R | N | R | N |
Aniline | N | – | – | R | – |
Diethylacetamide | N | L | N | R | N |
Dimethyl Formamide | N | R | N | R | N |
Pyridine | N | R | N | R | – |
Triethanolamine | – | R | – | R | N |
Group of Substance & Chemical Reagents |
Cellulose Acetate |
Nylon | PES | PTFE | PVDF |
MISCELLANEOUS | |||||
Formaldehyde Solution 30% | L | R | R | R | R |
Hydrogen Peroxide 30% | N | L | N | R | R |
Pyridine | N | R | N | R | R |
Silicone Oil & Mineral Oil | R | R | R | R | R |
R = | Recommended. No significant change observed in flow rate or bubble point of the membrane, nor visible indication of chemical attack. |
L = | Limited Recommended Use. Moderate changes in physical properties. The filter may be suitable for short term, noncritical use. |
N = | Not Recommended. The membrane may be unstable. |
– = | Insufficient Data. Information is not available. Trial testing is recommended. |
Top 9 Reasons to Use Restek Syringe Filters
- Protect any analytical system.
- Extend LC column lifetime.
- Achieve more reproducible analyses.
- Variety of membranes, porosities, and diameters available.
- Luer lock inlet provides strong, leak-tight syringe connection to withstand filtration pressure.
- Rugged polypropylene construction— autoclavable to 121 °C for 15 minutes (75 psi).
- Labeled (13, 25, and 30 mm, only) and color coded for easy identification.
- Convenient dispenser box.
- FREE sample pack available. Contact your Restek sales representative today.