Your web browser will no longer be supported by Restek.com as of 30 June 2021.
To avoid any interruption in access or functionality, install a current-generation web browser now. Learn more.

Versatile, High-Performing radiello Passive Air Samplers

  • 3D radial design outperforms 2D axial samplers.
  • Excellent low-level detection, even in short sampling times.
  • High capacity ensures accurate long-term sampling.
  • Easy to use—no pump required.
 

Now available from Restek, radiello passive diffusive air samplers feature a unique 3D radial design that provides several performance advancements compared to standard 2D flat axial air samplers. These samplers are a reliable, cost-effective alternative that offers both high sensitivity for low-level detection in short sampling periods, as well as higher capacity so accurate results can be obtained over longer sampling times. Versatile radiello air samplers are available for a wide range of chemical classes, so the choice of sampling material can be tailored to specific applications.

What Makes It Better?

The 3D cylindrical, coaxial design of radiello passive air samplers creates a large diffusive surface at a fixed, uniform distance from an adsorbent center column. Air enters from all sides and analytes are trapped on the large inner adsorptive surface. This results in a much higher sample uptake rate and capacity compared to standard 2D flat axial samplers.

The radiello Advantage

  • Higher sample uptake rate provides ppbv sensitivity in short sampling periods.
  • Higher capacity for more accurate long-term monitoring.
  • Easy to use—no pump required.
  • Not affected by face velocity.

Why Use radiello Air Samplers?

  • Unique 3D radial design is not prone to face velocity effects and delivers higher uptake rates than standard 2D flat axial devices.
  • Higher uptake rates give excellent low-level sensitivity, even for short sampling periods (down to 15 min).
  • Higher capacity allows you to sample accurately up to 30 days without bias from breakthrough or back diffusion.
  • Versatile, easy-to-use air sampler for a wide range of chemical classes is ideal for environmental monitoring, vapor intrusion, industrial hygiene, and personal air sampling.
  • Solvent and thermal desorption options available.
  • Features water repellant diffusive body and reusable holders.
  • Easy to use—no pump required.
 

Unique 3D Radial Design Means radiello Air Samplers Have a Much Higher Sample Uptake Rate than 2D Axial Samplers.

Compound 3D Radial radiello Sampler
Uptake Rate (mL/min)
2D Axial Sampler
Uptake Rate (mL/min)
Methanol 125 NS
Ethanol 102 44
Acetone 77 40
Benzene 80 36
Methyl ethyl ketone 79 36
Toluene 74 31
Ethylbenzene 68 27
n-Octane 53 27
Naphthalene 25 25
NS – Not suitable
 

Higher Sample Uptake Rates Mean High Sensitivity and Good Low-Level Detection, Even in Short Sampling Periods

The radiello Advantage

  • Increased response for all compounds compared to axial samplers.
  • Average increase in response of 163% for the 8 compounds shown below.
  • Excellent reproducibility with an average RSD of 5.6%.
 

figure-article-GNSS2960-02.jpg

radiello Samplers Have Higher Capacity than Axial Samplers, so You Can Report Accurate Results over Long Sampling Periods

In active sampling using a vacuum pump, the phenomenon of “breakthrough” occurs when the analyte band begins to move out of the adsorbent bed. If the concentration of analyte in the outlet air exceeds 10% of the concentration in the sampled air, any further pumping leads to analyte loss and an underrepresentation of the true environmental concentration. A similar situation, called “back diffusion,” occurs with diffusion sampling. Although, radiello air samplers can exhibit back diffusion, the high capacity of these devices allows much higher sample volumes to be processed before this effect is observed.

In this comparison, benzene is actively sampled (pumped) onto an activated charcoal adsorbent and also diffusively sampled using a radiello 130 sampler (both at 25 °C). For the pumped charcoal adsorbent (solid lines), breakthrough is reached at 35, 44, and 49 L when the benzene concentration is 100, 50, and 10 μg/m³, respectively. Whereas, for the radiello sampler (dotted lines) back diffusion does not occur until 1600, 2300, and 3050 L at the same concentrations.

figure-article-GNSS2960-03.jpg

Example Applications: radiello 145

The radiello 145 (RAD145) air sampler is designed specifically for airborne volatile organic compounds (VOCs). The 3D radial, diffusive design provides much higher uptake rates and more capacity than standard 2D flat axial, diffusive samplers. The higher sample uptake rate means that ppbv levels of VOCs can be detected in very short sampling times.

4-Hr Shipping Facility Air Sample with radiello 145

ColumnRtx-VMS, 60 m, 0.25 mm ID, 1.40 µm (cat.# 19916)
with MXT low-dead-volume connector (cat.# 20536)
Standard/SampleShipping facility air sample
Injectionon-column
Oven
Oven Temp.:40 °C (hold 7 min) to 250 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
28.8038226
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:6 mm ID
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Tune Type:BFB
Ionization Mode:EI
PreconcentratorMarkes Unity
  Trap 1 Settings
Type/Sorbent :radiello 145
Desorb temp.:350 °C
Desorb flow:50 mL/min
Desorb time:300 sec
  Trap 2 Settings
Type/Sorbent:Air Toxics
Cooling temp.:30 °C
Desorb temp.:310 °C
Desorb time:3 sec
InstrumentAgilent 7890B GC & 5977A MSD
Sample PreparationThe radiello 145 passive air sampler (RAD145) utilizes a stainless steel net cartridge packed with 350 mg of graphitized charcoal (Carbograph 4). Airborne volatile organic compounds (VOCs) were adsorbed to the charcoal and then thermally desorbed and analyzed by GC-MS.

Trap 1 conditions were used for radiello desorption. Trap 2 conditions were used for Unity desorption.

 

ColumnRtx-VMS, 60 m, 0.25 mm ID, 1.40 µm (cat.# 19916)
with MXT low-dead-volume connector (cat.# 20536)
Standard/SampleLaboratory air sample
Injectionon-column
Oven
Oven Temp.:40 °C (hold 7 min) to 250 °C at 30 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
28.8038226
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:6 mm ID
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Tune Type:BFB
Ionization Mode:EI
PreconcentratorMarkes Unity
  Trap 1 Settings
Type/Sorbent :radiello 145
Desorb temp.:350 °C
Desorb flow:50 mL/min
Desorb time:300 sec
  Trap 2 Settings
Type/Sorbent:Air Toxics
Cooling temp.:30 °C
Desorb temp.:310 °C
Desorb time:3 sec
InstrumentAgilent 7890B GC & 5977A MSD
Sample PreparationThe radiello 145 passive air sampler (RAD145) utilizes a stainless steel net cartridge packed with 350 mg of graphitized charcoal (Carbograph 4). Airborne volatile organic compounds (VOCs) were adsorbed to the charcoal and then thermally desorbed and analyzed by GC-MS.

Trap 1 conditions were used for radiello desorption. Trap 2 conditions were used for Unity desorption.

Example Applications: radiello 165

Monitoring airborne carbonyls (aldehydes and ketones) is important for protecting human and environmental health. A radiello 165 (RAD165) passive sampler is ideal for collecting these compounds from either indoor or outdoor air. The 3D radial design ensures high sample uptake rates, allowing ppbv level detection. The large capacity allows for longer sampling periods to be used when desired.

16-Hr Laboratory Air Sample with radiello 165

LC_AR0508
ColumnRaptor C18 (cat.# 9304A65)
Dimensions:150 mm x 4.6 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Temp.:30 °C
Standard/SampleLaboratory air sample
Diluent:2 mL acetonitrile extract
Inj. Vol.:5 µL
Mobile Phase
A:Water
B:Methanol:acetonitrile (650:50)*
Time (min)Flow (mL/min)%A%B
0.000.83070
5.000.82575
11.000.81090
11.010.80100
12.000.80100
12.010.83070
14.000.83070
DetectorUV/Vis @ 365, 4.8 nm
Preconcentratorradiello 165
InstrumentWaters ACQUITY UPLC H-Class
NotesThe radiello 165 passive air sampler (RAD165) utilizes a stainless steel net cartridge filled with 2,4-dinitrophenylhydrazine (DNPH) coated Florisil adsorbent. Airborne carbonyls react with the DNPH to form stable hydrazones. The derivatives were subsequently extracted for 30 minutes with 2 mL of acetonitrile. The extract was then analyzed by LC-UV.

*Mobile phase B was prepared by combining 650 mL methanol and 50 mL acetonitrile.
GNSS2960-UNV

Related Products

radiello Passive Diffusive Air Samplers

View Product