USLC Columns—Ultra Selective Liquid Chromatography

USLC Columns—Ultra Selective Liquid Chromatography

uslc_002

Choose Columns Fast. Develop Methods Faster.

  • Represents the widest range of reversed phase selectivity available, with just 4 stationary phases: Aqueous C18, IBD, Biphenyl, and PFP Propyl.
  • Simplifies column choice for fast, effective method development.
  • Offers greater control to more easily resolve closely eluting peaks.
  • Maximizes column screening efficiency by ensuring orthogonality between analytical runs.
  • Fits universally on any HPLC or UHPLC instrumentation.

Selectivity is the most influential factor affecting peak separation, or resolution. Restek Ultra Selective Liquid Chromatography (USLC) technology is the directed application of orthogonal selectivity to help the practicing chromatographer develop methods more efficiently and more effectively. The Restek USLC column set offers the widest range of selectivity in the industry using just 4 unique stationary phases—Aqueous C18, IBD, Biphenyl, and PFP Propyl—to simplify phase selection.

Restek USLC columns are available on two silica lines to accommodate any instrument platform or chromatographic application. Pinnacle DB columns (1.9 µm) feature base deactivated silica and are recommended for scalable HPLC and UHPLC analyses. Ultra columns (3 and 5 µm) use highest purity silica and are well suited for traditional HPLC analyses.

Pinnacle DB Biphenyl Columns (USP L11)

Stationary Phase Category: phenyl (L11)
Ligand Type: unique Biphenyl
Particle: 1.9 µm, 3 µm, or 5 µm, spherical
Pore Size: 140 Å
Carbon Load: 8%
End-Cap: yes
Surface Area: 150 m2/g
pH Range: 2.5 to 8
Maximum Temperature: 80 °C

Ultra Biphenyl Columns (USP L11)

Stationary Phase Category: phenyl (L11)
Ligand Type: unique Biphenyl
Particle: 3 µm or 5 µm, spherical
Pore Size: 100 Å
Carbon Load: 15%
End-Cap: yes
Surface Area: 300 m2/g
pH Range: 2.5 to 8
Maximum Temperature: 80 °C

Pinnacle DB PFP Propyl Columns (USP L43)

Stationary Phase Category: fluorophenyl propyl (L43)
Ligand Type: pentafluorophenyl propyl
Particle: 1.9 µm, 3 µm, or 5 µm, spherical
Pore Size: 140 Å
Carbon Load: 6%
End-Cap: yes
Surface Area: 150 m2/g
pH Range: 2.5 to 8
Maximum Temperature: 80 °C

Ultra PFP Propyl Columns (USP L43)

Stationary Phase Category: fluorophenyl propyl (L43)
Ligand Type: pentafluorophenyl propyl
Particle: 3 µm or 5 µm, spherical
Pore Size: 100 Å
Carbon Load: 11%
End-Cap: yes
Surface Area: 300 m2/g
pH Range: 2.5 to 8
Maximum Temperature: 80 °C

Pinnacle DB Aqueous C18 Columns (USP L1)

Stationary Phase Category: modified C18 (L1)
Ligand Type: proprietary polar modified and functionally bonded C18
Particle: 1.9 µm, 3 µm, or 5 µm, spherical
Pore Size: 140 Å
Carbon Load: 6%
End-Cap: no
Surface Area: 150 m2/g
pH Range: 2.5 to 8
Maximum Temperature: 80 °C

Ultra Aqueous C18 Columns (USP L1)

Stationary Phase Category: modified C18 (L1)
Ligand Type: proprietary polar modified and functionally bonded C18
Particle: 3 µm or 5 µm, spherical
Pore Size: 100 Å
Carbon Load: 15%
End-Cap: no
Surface Area: 300 m2/g
pH Range: 2.5 to 8
Maximum Temperature: 80 °C

Pinnacle DB IBD UHPLC Columns (USP L68)

Stationary Phase Category: polar-embedded alkyl (L68)
Ligand Type: proprietary polar functional embedded alkyl
Particle: 1.9 µm, spherical
Pore Size: 140 Å
Carbon Load: proprietary
End-Cap: no
Surface Area: 150 m2/g
pH Range: 2.5 to 8
Maximum Temperature: 80 °C

Ultra IBD Columns (USP L68)

Specialized Columns for Mixed Polar and Nonpolar Compounds

Stationary Phase Category: polar-embedded alkyl (L68)
Ligand Type: proprietary polar functional embedded alkyl
Particle: 3 µm or 5 µm, spherical
Pore Size: 100 Å
Carbon Load: 12%
End-Cap: no
Surface Area: 300 m2/g
pH Range: 2.5 to 8
Maximum Temperature: 80 °C

Technical Articles

Clinical, Forensic & Toxicology

USLC Columns Put the Right Tools in Your LC Method Development Toolbox - Understanding Selectivity in Reversed Phase Separations – A Simplified Approach to HPLC and UHPLC Column Selection

Column selectivity has the most significant influence on chromatographic peak separation, or resolution, so choosing the right column can greatly speed up HPLC and UHPLC method development. In this article, we discuss column choice and identify a set of just 4 stationary phases—Restek’s USLC column set—that encompasses the widest range of reversed phase selectivity available today.

Pharmaceutical

Column Choice: A Critical Factor for Successful UHPLC Integration

Column choice is a critical factor in successfully transferring methods between UHPLC and HPLC. Here, we discuss the column qualities that contribute to the successful integration of UHPLC technology.

Pharmaceutical

Novel Column Chemistry—High Impact, Low Cost Technology

Novel column chemistries are a simple change in an already budgeted consumable that can lead to optimized and more reliable methods—giving a fast return on a minimal investment.

Foods, Flavors & Fragrances

Comprehensive Pesticide Residue Analysis by LC-MS/MS Using an Ultra Aqueous C18 Column

Analysis of pesticide residues in foods is complicated by the large number of target compounds. LC-MS/MS analysis using an Aqueous C18 column provides more comprehensive results than GC alone.

General Interest

Tech Tip — Why do I Lose Retention on my C18 in Highly Aqueous Mobile Phases?

Learn what de-wetting is, and how to avoid it, by using an Aqueous C18 column, instead of a conventional C18, when analyzing polar compounds by HPLC.

PDF Document

USLC Columns: Choose Columns Fast. Develop Methods Faster.

Restek USLC columns offer the widest range of selectivity available and are an integral part of successful method development. Ideal for column switching systems, these columns provide orthogonal separations to create optimal resolution and robust methods—all in a 4-column set.

PDF Document

USLC Column Selection & Mobile Phase Adjustment Guide

USLC columns and this guide will help you easily choose the right stationary phase to target nearly any analyte in reversed phase or HILIC method development. It will also help you adjust your mobile phase to further improve results without guesswork or wasted time.

Related Links

Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us